Technical Data

common info
beam tube SR-9, channel B
distance from floor 1.00 m
max. Q-value 0.8/Å < Q < 2.9/Å
cross section of beam tube nose 140 x 90 mm2 (w*h)
maximum usable divergence 1° horizontal, 2° vertical

collimators
horizontal divergence 15', 30', 60'

monochromator
distance hot source to monochromator 6,0 m
cross section of beam at monochromator 40 x 100 mm2
Mechanic vertikal focussing (variable)
crystals Cu-(220), Cu-(420), Ge-(311) (available at summer 2005)
reflecting area 120 x 100 mm² (w*h) (seven horizontal elements)
monochromator 20°, 40°, 50°
range of wavelength 0.3 Å < λ < 1.1 Å

diffractometer
distance monochromator-sample 2.0 m (variable)
sample desk base on air cushions
capacity ca. 10 kN, diameter 500 mm
goniometer with detector holder, diameter 500 mm
ω goniometer, diameter 400 mm
acentric Euler cradle, diameter 400 mm (inner value)
double gonimeter for cryomagnet (optional)
sample -40° bis +120°

sample environment
air cooled furnace Tmax=1000°C
closed cycle cryostat Tmin=4 K

analyser
crystal PG-(002), area 150x30 mm² (w*h), thickness 2 mm

detector
distance sample-detector 0,35 m - 0.6 m
single counter 3He-counter tube, optimized for short wavelengths
2D-detector sensitive area 200*200 mm², available at summer 2005