ANTARES

Radiographie und Tomographie mit kalten Neutronen

Technical Data

Collimation and flux at the sample position

  • L/D = 200, 4 · 108 n/cm2s
  • L/D = 400, 1 · 108 n/cm2s
  • L/D = 800, 2.6 · 107 n/cm2s
  • L/D = 8000, 2.6 · 105 n/cm2s
  • Beam size up to 35 × 35 cm2

Neutron beam optics (optional)

  • Double Crystal Monochromator:
    2.7 Å ≤ λ ≤ 6.5 Å (1 %<Δ λ / λ <3 %)
  • Neutron Velocity Selector:
    2.5 Å ≤ λ≤ 8 Å (Δ λ / λ = 10 %)
  • Beam Filters:
    Cd filter for epithermal imaging
    Be filter to suppress wavelengths λ < 4 Å
    Sapphire filter to suppress fast neutrons

Sample table

XY-Phi-table:

  • Capacity: 500 kg
  • Travel: x = 800 mm, y = 600 mm
  • Rotation table: 360° rotation

Detection systems

  • Camera box with mirror and scintillation screens of different sizes from 6 x 6 cm² to 40 x 40 cm², screen thickness from 10 μm to 200 μm, plus X-ray screens
  • Standard detector: ANDOR cooled CCD camera, 2048 x 2048 pixels, 16 bit
  • Intensified triggerable iStar ANDOR cooled CCD camera, 1024 x 1024 pixels, 16 bit
  • Intensified NTSC video camera (30 fps) with analog frame grabber, MPEG-2 and DivX recording
  • DürrDental Image Plate scanner for arbitrary imaging plates, focus size 12.5 – 100 μm
  • Fuji BAS 2500 Image Plate scanner, focus size 25 – 100 μm
  • X-ray and neutron imaging plates
  • MAR345 image plate detector, 345 mm diameter, N-sensitive image plate

Dr. Burkhard Schillinger

Telefon: +49 89 289- 12185 / 14815

E-Mail: burkhard.schillinger@frm2.tum.de

Dr. Michael Schulz

Telefon: +49 89 289- 14718

E-Mail: michael.schulz@frm2.tum.de

ANTARES

Telefon: +49 89 289- 14815

E-Mail: